Gasmet On-line Series includes multicomponent gas analyzers for continuous monitoring applications. The Gasmet CX4015 incorporates a Fourier transform infrared spectrometer, a temperature controlled sample cell, and signal processing electronics. The analyzer is fully equipped for fixed installations and it offers versatility and high performance for all industrial users.

The Gasmet CX4015 is designed workplace air monitoring, and quality control applications. It is an ideal tool to measure components of interest in ambient conditions. The sample cell can be heated up to 50 °C. Sample cell absorption path length is selected according to the application.

The Gasmet CX4015 allows simple calibration using only single component calibration gases. The user can easily configure the analyzer for a new set of compounds.

Gasmet CX4015 can be integral part of Gasmet Continuous emissions monitoring system (CEMS II) or it can be operated as an individual analyzer.

AREAS OF APPLICATION : Occupational hygiene, Quality control.

Features


  • Highest possible sensitivity for low detection limits

Technical


General Parameters
Measuring principle
Fourier transform infrared, FTIR
Performance
Simultaneous analysis of up to 50 gas compounds
Response time, T90
Typically < 120 s, depending on the gas flow and measurement time
Operating temperature
Long term 5 – 30 °C Non-condensing air conditioning recommended
Storage temperature
-20 – 60 °C, non condensing
Power supply
100-115 or 230 V / 50-60 Hz
Power consumption
Average 150 W, maximum 300 W
Spectrometer
Resolution
8 cm-1 or 4 cm-1
Scan frequency
10 scans / s
Detector
Peltier cooled MCT
Source
SiC, 1550 K
Beamsplitter
ZnSe
Window material
ZnSe
Wave number range
900 - 4 200 cm-1
Sample Cell
Structure
Multi-pass, fixed path length 9.8 m
Material
100 % rhodium coated aluminum
Mirrors
Fixed, protected gold coating
Volume
0.4 liters
Connectors
Inlets Swagelok 6 mm Outlet Swagelok 8 mm
Gaskets
Viton® O-rings
Temperature
50 °C, maximum
Valve
Manual Swagelok
Window material
AR coated ZnSe
Measuring parameters
Zero point calibration
24 hours, calibration with nitrogen (5.0 or higher N2 recommended)
Zero point drift
< 2 % of measuring range per zero point calibration interval
Sensitivity drift
None
Linearity deviation
< 2 % of measuring range
Temperature drifts
< 2 % of measuring range per 10 K temperature change
Pressure influence
1 % change of measuring value for 1 % sample pressure change. Ambient pressure changes measured and compensated.
Electrical connectors
Digital Interface
9-pole D-connector for RS-232. Analyzer is connected to an external computer via RS-232C cable. The external computer controls Gasmet.
Power connection
Standard plug CEE-22
Gas inlet and outlet conditions
Gas temperature
Non-condensing, the sample gas temperature should be the same as the sample cell temperature
Flow rate
120 - 600 liters per hour
Gas filtration
Filtration of particulates (2 µ) required
Sample gas pressure
Ambient
Sample pump
External, not included
Electronics
A/D converter
Dynamic range 95 dB,
Signal processor
32-bit floating point DSP, 120 MFLOPS
Computer
External, not included
Analysis software (for external PC)
Operating system
Windows 7 (32-bit)
Analysis software
Calcmet for Windows
Options
Sample cell
Multi-pass, fixed path 2.5 m or 5.0 m
Pressure measurement
Inside sample cell
External PC
Gasmet PC for control and analysis with optional alarm relay boards and analog inputs or outputs
Sample cell gaskets
Kalrez®
Enclosure
Material
Aluminium
Dimensions (mm)
482 * 196 * 450
Weight
17 kg
CE label
According to EMI guideline 89/336/EC®